IMAGE MAGNIFICATION Sometimes the distance between test specimen and image detector is increased to obtain magnification in the image. Magnification is especially useful when parts being inspected and their details are very small. The farther the test specimen is from the image detector, the greater the magnification achieved. The amount of magnification can be calculated using the following formula. M = (a + b) / a
Figure 2. Sketch showing geometric variables that affect magnification.
In RTR, focal-spot sizes as small as 10 microns are
used to allow for magnification of the image and yet keep the penumbra
to a minimum. One micron or micrometer (my-króh-mee-tuhr)
equals one-millionth of a meter (0.000001 meter, or .001 mm, or .00003937
inches, or 1/25,400 inch). If a newspaper sheet is .003 inches thick,
it is approximately 76 microns thick. One disadvantage of magnification
is that the field of view is decreased. When the viewable area becomes
smaller, this generally increases the inspection time. Refer to the
applications booklet for a worksheet involving magnification problems.